Alarm codeO2
The O2 sensor may be new and needs calibration, or its readings are inaccurate due to calibration issues or depletion. The manual's first action is: “Calibrate the O2 sensor.”
- Model
- Hamilton Medical HAMILTON-C6
- Display shows
- O2
- Type
- Alarm code
- Source
- Manual p. 219
What O2 means
Low priority. O2 sensor calibration data is not within expected range, or sensor is new and requires calibration. p. 219
Common causes
- Low priority. O2 sensor calibration data is not within expected range, or sensor is new and requires calibration. p. 219
- Low priority. O2 measurement accuracy is reduced. p. 219
- Low priority. The O2 sensor is depleted. p. 219
- Low priority. There is no signal from the O2 sensor. p. 219
- Low priority. The incorrect type of O2 sensor is installed. p. 219
What to do
- Calibrate the O2 sensor. p. 219
- Verify temperature settings are within environmental specifications. p. 219
- Replace O2 sensor if required. p. 219
- Have the ventilator serviced. p. 219
- If using a paramagnetic O2 sensor, calibrate the sensor or have the ventilator serviced. p. 219
- Replace O2 sensor. p. 219
- To ensure O2 is always monitored, replace O2 sensor ASAP or use external ISO 80601-2-55-compliant O2 monitor. p. 219
- Ensure a Hamilton Medical O2 sensor is used and it is properly installed. p. 219
Before you service the product, read the safety instructions in the manual p. 32.
If O2 is still not resolved
This page summarises what the manufacturer’s manual says about O2; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Hamilton Medical. Found a mistake? Report an error.