NOx calibration abnormality occurs when zero gas is not flowing, the gas chamber is contaminated causing zero offset, or the detector sensitivity has decreased. The manual's first action is: “ENT: Continue calibration”
- Model
- Fuji Electric ZRE INZ-TN1ZREb-C
- Display shows
- NOx calibration abnormality (NO.5)
- Type
- Error code
- Source
- Manual p. 70
What NO.5 means
NOx calibration abnormality p. 70
Common causes
- Zero gas not flowing p. 70
- Gas chamber contamination causing zero offset p. 70
- Detector sensitivity decreased p. 70
What to do
If NO.5 is still not resolved
The manual gives no further instruction for this code. Open the code table in the PDF or contact Fuji Electric service.
This page summarises what the manufacturer’s manual says about NO.5; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Fuji Electric. Found a mistake? Report an error.