Error codeE11
The balance cannot stabilize due to an environmental problem. The manual's first action is: “Prevent vibration, drafts, temperature changes, static electricity and magnetic fields.”
- Model
- A&D GF series
- Display shows
- Stability error (E11)
- Type
- Error code
- Source
- Manual p. 71
What E11 means
Stability error p. 71
Common causes
- The balance can not stabilize due to an environmental problem. p. 71
What to do
- Prevent vibration, drafts, temperature changes, static electricity and magnetic fields. p. 71
- Refer to "3. PRECAUTIONS" for details on the operating environment and "7. RESPONSE ADJUSTMENT" about adapting the balance to the environment. p. 71
- To return to the weighing mode, press the CAL key. p. 71
- Prevent vibration, drafts, temperature changes, static electricity and magnetic fields. Refer to "3. PRECAUTIONS" for details on the operating environment and "7. RESPONSE ADJUSTMENT" about adapting the balance to the environment. To return to the weighing mode, press the CAL key. p. 71
If E11 is still not resolved
The manual gives no further instruction for this code. Open the code table in the PDF or contact A&D service.
This page summarises what the manufacturer’s manual says about E11; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with A&D. Found a mistake? Report an error.