CONOX Consciousness and Nociception Monitor Installation Guide
This document provides a standardized procedure for installing and operating the CONOX monitor for consciousness and nociception monitoring. It details required materials, roles for anesthesiologists, nurses, and technicians, equipment setup, skin preparation, sensor fixation locations, and impedance verification.
The guide also explains the monitor's main screen indices (qCON, qNOX, BSR, EMG, SQI) with ideal ranges, basic interpretation guidelines, Bluetooth connectivity for data export via the CONOX VIEW Android app, accessing stored monitoring sessions, and safe sensor removal and storage, including a critical warning to disconnect sensors during defibrillation. This manual is written in Portuguese.
Covers: Objectives and Required Material · Roles and Responsibilities · Equipment Preparation · Skin Preparation and Sensor Placement · Equipment Operation · Impedance Verification
Latest CONOX manuals
CONOX Consciousness and Nociception Monitor Installation Guide
Installation Guide 6 pages · 1.1 MB · PTHospital protocol for installing and using the CONOX consciousness and nociception monitor, covering sensor placement, impedance checks, index…
Which document do I need?
Installation Guide
Mounting, wiring and first power-up. Read it before the hardware goes in.
Questions answered in CONOX manuals
What is the ideal sensor impedance for the CONOX monitor?
The ideal impedance for all three electrodes is ≤5 kΩ; values between 6–10 kΩ require re-preparation, and above 10 kΩ triggers a 'TH' reading requiring mandatory re-fixation.
From: CONOX Consciousness and Nociception Monitor Installation Guide
Can the CONOX monitor connect to a smartphone for data export?
Yes, the monitor has Bluetooth and can pair with the CONOX VIEW app on Android devices using the equipment's serial number as the password.
From: CONOX Consciousness and Nociception Monitor Installation Guide