Ulke ULDW-5A Earth Ground Resistance Tester User Manual
The Ulke ULDW-5A is a ground resistance tester that uses a variable-frequency AC source and microcomputer control. The manual says it is intended to measure power-frequency grounding impedance, contact voltage, step voltage, equivalent power-frequency parameters and soil resistivity, and that it can measure without interrupting power.
The manual describes the instrument's principles, panel, measurement wiring and operating steps. The tester uses 45 Hz and 55 Hz for variable-frequency measurement and has a maximum output current of 5 A; its grounding impedance range is 0 to 200 Ω and its soil resistivity range is 0 to 1000 Ω·m. This manual is written in Chinese.
Covers: Instrument Overview · Performance Features · Technical Specifications · Instrument Principle · Panel Introduction · Measurement Wiring
Latest 优利克电力 manuals
Ulke ULDW-5A Earth Ground Resistance Tester User Manual
User Manual 15 pages · 652 KB · ZHUser manual for the Ulke ULDW-5A earth ground resistance tester, covering panel controls, measurement wiring, operation, specifications and troubleshooting.
Which document do I need?
User Manual
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in 优利克电力 manuals
What measurement frequencies does the ULDW-5A use?
In variable-frequency mode it uses 45 Hz and 55 Hz; in fixed-frequency mode it uses a strict 50 Hz.
From: Ulke ULDW-5A Earth Ground Resistance Tester User Manual
What is the maximum output current of the tester?
The maximum output current is AC 5 A, using 45 Hz and 55 Hz dual-frequency sine wave output.
From: Ulke ULDW-5A Earth Ground Resistance Tester User Manual