Questions answered in AJA International manuals

What should RF biasing of the substrate be limited to?

RF biasing of the substrate should be limited to 50 WATTS; 10-30 W should be more than adequate for substrates up to 6 “ diameter.

From: AJA International ATC & ATC Orion Series Sputtering Systems Installation & Operation Manual

What incoming cooling water temperature is recommended in humid environments?

In humid environments, the incoming cooling water temperature should be elevated to 25 °C to prevent condensation.

From: AJA International ATC & ATC Orion Series Sputtering Systems Installation & Operation Manual