SensorData HT-912-T-01 Reference Bottle Calibration Manual
The document provides an instruction manual for SensorData reference bottles (HT-912-T-01) used to check and calibrate relative humidity transmitters. It lists five bottle types with saturated salt solutions at 12%, 33%, 55%, 75%, and 90% RH, each with a specific type number. The manual includes lifetime guidelines, handling precautions, and temperature compensation notes.
Two calibration procedures are detailed: a standard 2-point calibration using 12% and 90% bottles with zero and span adjustments, and a quicker 1-point field calibration using either 75% or 90% bottle with zero adjustment only. Both require pre-stabilization and approximately 3-hour stabilization periods. References to NBS and PTB standards are cited.
Covers: Test Certificate · Checking Relative Humidity Transmitters · Life Time · 2-Point Calibration Procedure · 1-Point Calibration Procedure · Cautions and References
Latest SensorData manuals
SensorData HT-912-T-01 Reference Bottle Calibration Manual
User Manual 2 pages · 637 KB · ENInstruction manual for SensorData HT-912-T-01 reference bottles used to calibrate HVAC relative humidity probes via 1-point and 2-point procedures.
Which document do I need?
User Manual
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in SensorData manuals
How long should reference bottles be pre-stabilized before calibration?
Reference bottles must be pre-stabilized for a minimum of 1 hour at the present ambient temperature before use.
From: SensorData HT-912-T-01 Reference Bottle Calibration Manual
What is the expected lifetime of the calibration bottles?
The bottles have a lifetime of approximately 2 years depending on careful handling, provided the salt solution remains saturated with visible crystals and water.
From: SensorData HT-912-T-01 Reference Bottle Calibration Manual