Klay Instruments 8000 Series Pressure and Level Transmitters User Manual
This manual provides instructions for the Klay Instruments SERIES 8000 and SERIES 8000-SAN solid-state pressure and level transmitters. These devices utilize a bridge resistive silicon sensor with a stainless steel foot and flush diaphragm to measure pressure in industrial processes, converting the signal to a 4-20 mA output.
Covers: Introduction · Description Series 8000-SAN · Description Series 8000 · Dimensional Drawing · Installing Weld-on Nipple · Installing Transmitter
Latest KLAY-INSTRUMENTS manuals
Klay Instruments 8000 Series Pressure and Level Transmitters User Manual
User Manual 9 pages · 459 KB · ENUser manual for Klay Instruments 8000 and 8000-SAN solid-state pressure and level transmitters, covering installation, wiring, and calibration.
Which document do I need?
User Manual
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in KLAY-INSTRUMENTS manuals
What is the calibration sequence for the transmitter?
Set output to 4 mA, apply air pressure, set output to 20 mA, remove pressure, check output is 4 mA, install transmitter, and set output to 4 mA based on mounting position.
From: Klay Instruments 8000 Series Pressure and Level Transmitters User Manual
Can the transmitter be used in hazardous areas?
Yes, the SERIES 8000 and 8000-SAN are certified for applications in hazardous areas with specific ATEX and IECEx ratings.
From: Klay Instruments 8000 Series Pressure and Level Transmitters User Manual