YSI 550A Portable Dissolved Oxygen Meter User Manual
The YSI 550A is a portable dissolved oxygen instrument powered by four No. 3 alkaline batteries, offering approximately 2000 hours of operation on a full set. The device features a waterproof housing that should not be opened by unauthorized personnel to maintain the factory seal. It includes a convenient calibration/storage chamber with a sponge for creating a 100% water-saturated air environment for probe storage and calibration. This manual is written in Chinese.
Read onlineYSI 550A specifications
| Dissolved oxygen range (% saturation) | 0 to 500% air saturation |
|---|---|
| Dissolved oxygen range (mg/L) | 0 to 50 mg/L |
| Dissolved oxygen resolution (mg/L) | 0.01 mg/L or 0.1 mg/L; selectable |
| Dissolved oxygen accuracy (0 to 20 mg/L) | ±0.3 mg/L or 2% of reading, whichever is greater |
| Temperature range | -5 to +45°C (0.1°C resolution, ±0.3°C) |
| Power | 4 alkaline C cells |
| Battery life | Over 2000 hours |
| Size | 4.7 in. width; 9 in. length (11.9 cm x 22.9 cm) |
| Weight with batteries | 2 lb. (0.91 kg) |
| Water resistance | IP-67 |
| Cable lengths | 12, 25, 50 and 100 feet |
| Salinity input | Manual salinity input from 0 to 70 ppt |
| Warranty | Three years for the instrument, one year for the DO module |
Source: YSI 550A Dissolved Oxygen Instrument System Specifications
All YSI 550A documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
What dissolved oxygen range does the YSI 550A measure?
It measures dissolved oxygen from 0 to 500% air saturation and from 0 to 50 mg/L, with selectable resolution.
From: YSI 550A Dissolved Oxygen Instrument System Specifications
How long does the battery last?
The instrument runs on four alkaline C cells with over 2000 hours of battery life.
From: YSI 550A Dissolved Oxygen Instrument System Specifications
How long does a new set of alkaline batteries last in the YSI 550A?
A new set of alkaline batteries can power the instrument for approximately 2000 hours.
What is the required stirring speed when calibrating the YSI 550A in mg/L mode?
The probe must be stirred or agitated in the sample at a frequency of at least 1/2 foot per second.