The Thorlabs Blueline TED 200 is a precision temperature controller for laser diodes and detectors. It can stabilize wavelength, reduce detector noise, tune or modulate wavelength by regulating temperature. The front panel has an illuminated 4 1/2-digit LCD display, and the PID control loop shares can be set. It accepts thermistor, AD 590, or LM 335 sensors, and connects the temperature sensor and TEC element with a 9-pin D-sub plug.
The manual contains general description, safety, warranty, features, technical data, operating elements, connections, operation, maintenance, service, and listings. Safety instructions say to check the protective conductor, line voltage setting, and fuses before applying power. Maintenance and service sections cover care, cleaning, repair, line voltage, fuses, sensor calibration, and troubleshooting.
Thorlabs TED 200 specifications
| TEC Current Output Control Range | 0 to ±2 A |
|---|---|
| Compliance Voltage | >6 V |
| Maximum Output Power | 12 W |
| Current Measurement Resolution | 1 mA |
| Current Measurement Accuracy | ±10 mA |
| Noise and Ripple (typ.) | <1 mA |
| TEC Current Limit Setting Range | 0 to >2 A |
| Thermistor Sensor Control Range | 10 W to 20 kW |
| Temperature Stability (IC sensors) | <0.002 °C (24 h) |
| Temperature Accuracy (IC sensors) | ±0.1 °C |
Source: Thorlabs TED200C Thermoelectric Temperature Controller Operation Manual
All Thorlabs TED 200 documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
How can the TEC current limit be adjusted?
Turn the potentiometer labeled "LIM I" on the rear panel to set the TEC current limit.
From: Thorlabs TED200C Thermoelectric Temperature Controller Operation Manual
What indicates that over‑temperature protection is active?
The LED labeled "OTP" lights and a short warning beep is heard when over‑temperature protection engages.
From: Thorlabs TED200C Thermoelectric Temperature Controller Operation Manual
What temperature sensors can be used with the TED 200?
The TED 200 can use a thermistor, AD 590, or LM 335 temperature sensor.
From: Thorlabs Blueline Series Thermoelectric Temperature Controller TED 200 Operation Manual
What is the TEC output control range?
The TEC output control range is 0 to 2 A.
From: Thorlabs Blueline Series Thermoelectric Temperature Controller TED 200 Operation Manual