SKC 210-1000 Series Pocket Pump Operating Instructions
The SKC 210-1000 Series Pocket Pump is a low‑flow sample pump designed for industrial hygiene sampling of organic gases and vapors. It features a compact, lightweight design with computer‑compatible circuitry and optional ATEX certification.
This manual provides detailed guidance on pump setup, flow and pressure calibration, keypad operation, battery care, accessory options, and warranty information. Safety notices, performance profiles, and troubleshooting tips are included to help users operate the pump safely and effectively.
SKC 210-1000 Series specifications
| Flow Range (Constant Flow) | 20 to 225 ml/min |
|---|---|
| Flow Accuracy | ± 5% of set value |
| Operating Temperature (Sampling) | -7 to 43°C (20 to 110°F) |
| Charging Temperature | 5 to 38°C (40 to 100°F) |
| Battery Type | NiCd or NiMH rechargeable |
| Dimensions | 11.4 x 5.6 x 3.6 cm |
| Weight | 142 grams |
All SKC 210-1000 Series documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
What is the theoretical sampling time at 200 ml/min and 20 inches water column backpressure?
The theoretical sampling time is approximately 6 hours.
How does the battery life change if a NiMH battery is used instead of NiCd?
Using a NiMH battery can increase sampling time to approximately 12 hours in most environments.
How long can the Pocket Pump run on a NiMH battery at 200 mL/min flow rate?
With a NiMH battery the pump can run for up to about 12 hours at 200 mL/min and 10 inches water back‑pressure.