Mettler-Toledo 770MAX Manuals

ManualLib has 2 documents for the Mettler-Toledo 770MAX (Other Electronics): 2 user manuals. Every file can be read online or downloaded as a PDF for free.

User Manual 2

Key facts

Model
770MAX
Industry
Industrial
Documents
2
Languages
EN · ZH
Last added
2026-10-23

Start here

Mettler-Toledo 770MAX Multiparameter Analyzer Instruction Manual

Mettler-Toledo 770MAX Multiparameter Analyzer Instruction Manual

User Manual 62 pages · 436 KB · EN

The Mettler-Toledo 770MAX is a multiparameter analyzer and transmitter designed for high-accuracy measurement and control. It accepts inputs from up to four Smart Sensors plus two additional pulse flow sensors. Supported parameters include resistivity, conductivity, pH, ORP, dissolved oxygen, dissolved ozone, TOC, flow, temperature, and pressure. The device supports derived measurements such as total dissolved solids, % rejection, and calculated pH. It features standard outputs including four isolated analog signals, RS232, and two TTL discrete output ports, with optional additional analog signals and relays available.

Read online

Mettler-Toledo 770MAX specifications

Functional Sensor inputs4 Smart Sensor channels, 2 pulse flow channels
Conductivity Range (Constant 2-E Cell)0.01 µS/cm to 2000 µS/cm
Resistivity Range (Constant 2-E Cell)0.5 k Ω-cm to 1000 M Ω-cm
pH Range-1 to 15 pH
ORP Range-1500 to +1500 mV
Temperature Range-40° to 200°C
Dissolved Oxygen Range0-15,000 ppb
Analog Output SignalsFour standard, four additional--optional
Serial OutputRS232 standard

Source: Mettler-Toledo 770MAX Multiparameter Analyzer Instruction Manual

All Mettler-Toledo 770MAX documents by type

User Manual 2

Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.

Questions answered in these manuals

What sensors are compatible with the 770MAX?

The 770MAX accepts inputs from up to four Smart Sensors plus two additional pulse flow sensors.

From: Mettler-Toledo 770MAX Multiparameter Analyzer Instruction Manual

Can Smart Sensors from the previous generation 770PC be used with the 770MAX?

No, Smart Sensors used with the previous generation 770PC instrument are not compatible with 770MAX due to signal processing and connector differences.

From: Mettler-Toledo 770MAX Multiparameter Analyzer Instruction Manual

What power supply voltage and frequency does the 770MAX require?

The analyzer operates from 100‑240 VAC at 47‑63 Hz.

From: Mettler-Toledo 770MAX Multi-Parameter Analyzer User Manual

How many analog output channels are available?

It provides a 4‑20 mA RS232 output and two TTL digital outputs.

From: Mettler-Toledo 770MAX Multi-Parameter Analyzer User Manual