Key facts
- Brand
- ifm
- Model
- PI27XX
- Documents
- 2
- Languages
- PL
- Last added
- 2011-11-19
Start here

IFM PI27xx Series Pressure Sensor User Manual
This manual provides comprehensive instructions for the IFM PI27xx electronic pressure sensor. It includes safety guidelines, description of functions, installation steps, wiring, menu navigation, parameter settings, diagnostic features and detailed technical specifications.
The document is organized into sections such as safety instructions, sensor functions, mounting, electrical connection, menu structure, parameter configuration, operation, dimensional drawing and technical data. It also contains factory‑default settings and calibration procedures for adapting the sensor to specific applications. This manual is written in Polish.
ifm PI27XX specifications
| Supply voltage | 18‑32 V DC |
|---|---|
| Current consumption | < 45 mA |
| Nominal current | 250 mA |
| Voltage drop | < 2 V |
| Power‑on delay | 0.4 s |
| Switching output response time | 3 ms |
| Switching frequency | 125 Hz |
| Analog output range | 4‑20 mA / 20‑4 mA |
| Maximum load resistance | (UB‑10) × 50 Ω |
All ifm PI27XX documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
How can I restore the sensor to factory settings?
Use the menu option "Przywrócenie ustawień fabrycznych" to reset the device to its factory defaults.
What is the supply voltage range for the PI27xx sensor?
The sensor operates from 18 V to 32 V DC.
How can the sensor be calibrated?
Calibration is performed by defining two calibration points (CP1 and CP2) and adjusting the zero‑point (COF) as described in the calibration section.
What is the default pressure measurement type?
The device measures relative pressure.