Key facts
- Brand
- Hioki
- Model
- 3197
- Category
- Analyzers
- Industry
- Electronics
- Documents
- 2
- Languages
- ZH · EN
- Last added
- 2026-11-09
Start here

HIOKI 3197 User Manual
The HIOKI 3197 Power Quality Analyzer is a portable instrument for monitoring voltage and current disturbances such as inrush current, swells, dips, transients, and interruptions. It records power, power factor, active/reactive energy, demand, and load changes, displaying data as graphs or numerical values.
The manual explains QuickSet automatic configuration, wiring verification with vector maps, event detection settings, data recording options, and PC connectivity via USB. It also lists specifications, environmental limits, and optional clamp‑on sensors for various current ranges.
Hioki 3197 specifications
| Display | 4.7-inch color STN LCD |
|---|---|
| Display languages | Chinese (Simplified), Japanese, English, Korean |
| Memory capacity | 4 MB (battery backup; can be divided into 4 parts) |
| Maximum recording time | 125 days |
| Battery life | About 6 hours |
| Dimensions and weight | 128 W x 246 H x 63 D mm, 1.2 kg |
| External interface | USB (Ver. 2.0) |
| Maximum input voltage | Voltage port AC 780 V (peak 1103 V); current port AC 1.7 V (peak 2.4 V) |
| Voltage measurement range | 600.0 V |
| Frequency measurement range | 45.00 to 66.00 Hz |
| Operating temperature and humidity | 0 to 40 C, below 80% RH (no condensation) |
| Measurement category | CAT III 600 V; CAT IV 300 V |
All Hioki 3197 documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
What power abnormalities can the Hioki 3197 detect?
It detects voltage surges (rise), voltage sags (drop), momentary power failures, inrush current, and instantaneous overvoltage.
What circuits can the Hioki 3197 measure?
It supports single-phase 2-wire, single-phase 3-wire, three-phase 3-wire (3P3W2M, 3P3W3M), and three-phase 4-wire (3P4W, 3P4W2.5E) circuits.
What is the maximum amount of data the internal memory can store?
The internal memory can hold up to 4 MB, allowing a maximum recording time of 125 days.
From: HIOKI 3197 User Manual
Which interface is used to transfer data to a PC?
Data can be transferred via a USB 2.0 (Full Speed) connection to a Windows 2000/XP computer.
From: HIOKI 3197 User Manual