Hach LDOTM HQ10 Portable Dissolved Oxygen Meter User Manual
This manual provides step-by-step instructions for operating the Hach LDOTM HQ10 portable dissolved oxygen meter. It covers setup, calibration procedures, pH testing (for HQ20 models), data storage and retrieval, and troubleshooting. The device supports measurements from 0 to 20 mg/L oxygen with ±0.1 mg/L accuracy in the 0–8 mg/L range. It also includes safety warnings and technical specifications for proper use and maintenance. This manual is written in Chinese.
Read onlineHach LDOTMHQ10 specifications
| Oxygen Measurement Range | 0~20 mg/L (ppm), 0~200% saturation |
|---|---|
| Accuracy | ±0.1 mg/L for 0~8 mg/L; ±0.2 mg/L for >8 mg/L |
| Temperature Range | 0~50℃ (32~122℉) |
| Resolution | Oxygen: 0.01 or 0.1 mg/L; % Saturation: 0.1% |
| pH Range (HQ20 only) | -2.00~19.99 |
| pH Resolution (HQ20 only) | 0.001/0.01/0.1 pH units |
| Power Requirements | 4 AA alkaline batteries or external power via power adapter |
| Environmental Conditions | 0~50℃ (32~122℉), 90% relative humidity, no condensation |
| Dimensions | 21.2×8.7×4.2 cm (8.35×3.43×1.65 in) |
| IP Rating | IP67 for instrument, IP40 for power adapter |
Source: Hach LDOTM HQ10 Portable Dissolved Oxygen Meter User Manual
All Hach LDOTMHQ10 documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
What is the oxygen measurement range of the LDOTM HQ10?
The oxygen measurement range is 0 to 20 mg/L (ppm) or 0 to 200% saturation.
From: Hach LDOTM HQ10 Portable Dissolved Oxygen Meter User Manual
What is the accuracy of the dissolved oxygen measurement?
The accuracy is ±0.1 mg/L for measurements between 0 and 8 mg/L, and ±0.2 mg/L for measurements above 8 mg/L.
From: Hach LDOTM HQ10 Portable Dissolved Oxygen Meter User Manual
What is the measurement range for dissolved oxygen?
The dissolved oxygen meter measures a range of 0 to 20 mg/L (ppm).
From: Hach LDOTMHQ10/LDOTMHQ20 Portable Dissolved Oxygen Meter User Manual
What is the accuracy of the dissolved oxygen measurement?
The accuracy is ±0.1 mg/L for readings between 0 and 8 mg/L.
From: Hach LDOTMHQ10/LDOTMHQ20 Portable Dissolved Oxygen Meter User Manual
What safety warnings are included in the manual?
The manual uses DANGER, CAUTION and NOTE symbols to highlight safety information and required precautions.
From: Hach LDOTMHQ10 and LDOTMHQ20 pH and Dissolved Oxygen Meter User Manual
What is the measurement range for dissolved oxygen?
The device measures dissolved oxygen from 0 to 20 mg/L.
From: Hach LDOTMHQ10 and LDOTMHQ20 pH and Dissolved Oxygen Meter User Manual