Gaymar MTA-4700 Manuals

ManualLib has 2 documents for the Gaymar MTA-4700: 1 service manual, 1 user manual. Every file can be read online or downloaded as a PDF for free.

Service Manual 1User Manual 1

Key facts

Brand
Gaymar
Model
MTA-4700
Documents
2
Languages
EN
Last added
2023-10-18

Start here

Gaymar MTA-4700 Meditherm Operating Instruction Manual

Gaymar MTA-4700 Meditherm Operating Instruction Manual

User Manual 4 pages · 1.5 MB · EN

The Gaymar MEDI-THERM System is designed to regulate patient temperature by circulating temperature-controlled water through hyperhypothermia blankets. The Model MTA-4700 machine can be operated in two basic modes: Automatic Mode, which regulates patient temperature to a set point, and Manual Mode, where the operator adjusts blanket water temperature.

Read online

Gaymar MTA-4700 specifications

Warranty Periodone (1) year
Compressor Warrantyfour (4) year pro-rated
Water Requirementdistilled water only

Source: Gaymar Medi-Therm Hyper/Hypothermia Service Manual

All Gaymar MTA-4700 documents by type

Service Manual 1

Disassembly, diagnostics and part numbers, for repair.

User Manual 1

Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.

Questions answered in these manuals

What should be checked every 20 minutes during patient therapy?

The patient and blanket water temperature must be checked every 20 minutes or as often as directed by the physician.

From: Gaymar Medi-Therm Hyper/Hypothermia Service Manual

What happens if power is interrupted during therapy?

The unit will go into standby mode (no therapy) and require resetting by the user before therapy can be continued.

From: Gaymar Medi-Therm Hyper/Hypothermia Service Manual

What operating modes does the MTA-4700 support?

The unit supports Automatic Mode, Manual Mode, and Patient Temperature Monitoring mode.

From: Gaymar MTA-4700 Meditherm Operating Instruction Manual

What type of probe is compatible with this unit?

The unit accepts a YSI 400 Series Probe or equivalent.

From: Gaymar MTA-4700 Meditherm Operating Instruction Manual