Chinese user manual for the Yingweite HVSN high-voltage point selection module: 32 points per module, up to 128 points per host, programming commands and…
The Yingweite HVSN is a high-voltage point selection module used together with an external insulation resistance meter that has a communication interface, such as the Tonghui TH2683A, for high-voltage insulation resistance or high-voltage measurement. Under host control, the user can switch any combination of points to the positive or negative test terminal of the meter, and parallel connection of several points allows fast multi-point insulation resistance measurement.
One HVSN module has 32 points: points 0#–15# are fixed to the positive test terminal of the meter, while 16#–31# can be wired to the positive or negative terminal as required. A single host can be expanded with up to 4 selection modules, for 128 points in total. The document covers the overview, working principle, programming method with example command strings for the A, B, C and D layers, precautions, and service and repair information. This manual is written in Chinese.
| Points per module | 32 |
|---|---|
| Points fixed to positive test terminal | 0#-15# |
| Points selectable to positive or negative terminal | 16#-31# |
| Maximum expansion per host | 4 modules, 128 points |
| Communication interface | RS232 |
| Maximum relays driven from host power | 6 |
| External drive power supply | 12V/2A |
One module has 32 points. Points 0#–15# are fixed to the positive test terminal of the insulation resistance meter, and 16#–31# can be wired to the positive or negative terminal as needed.
A single host can be expanded with up to 4 selection modules, giving 128 points in total.
Yes. When the two are mixed, the HVSN high-voltage point selection module should be installed below the P32SN point selection module.
An external 12V/2A drive power supply must be connected to V+ and GND, and the shorting cap on the JUMP socket must be removed.