This manual provides operating instructions, safety guidelines, and technical specifications for the Yiheng LHH-SD Series Pharmaceutical Stability Test Chamber.
The Yiheng LHH-SD Series Pharmaceutical Stability Test Chamber is a laboratory device designed for environmental science, genetics, and biological research. It features a vertical frame structure with a mirror stainless steel inner chamber, adjustable shelves, and insulation made of polyester foam. The unit utilizes a Swiss Rohde & Schwarz humidity and temperature transmitter for high precision and includes a microcomputer controller for stable operation. This manual is written in Chinese.
| Temperature Control Range | 0°C ~ 65°C |
|---|---|
| Humidity Control Range | 40% ~ 95% RH |
| Temperature Fluctuation | ±0.5°C |
| Humidity Fluctuation | ±3.0% RH |
| Temperature Uniformity | ±2.0°C |
| Power Supply | AC220V±10% 50Hz |
The chamber must be placed on a hard, flat surface in a horizontal state, with a clearance of at least 900mm in front, 500mm at the back, and 300mm on the sides and top.
If the actual temperature exceeds the limit setting, the independent temperature limit alarm system automatically interrupts the heating system to ensure safety.