User guide for the Vishay Transducers VT100 weight indicator covering operation, installation, configuration, calibration, and print formats.
The VT100 Weight Indicator User's Guide covers operation, installation, configuration, calibration, serial communication, and print formats for the VT100 weight indicator. It includes safety instructions, warranty information, a declaration of conformity, and appendices with commands, conversion factors, ASCII information, error messages, and display characters.
The guide describes front-panel and configuration-mode operations, test and panel modes, and electronic data processing commands. Detailed specifications are listed in the Specifications section.
| Storage temperature | -10°C to +70°C (14°F to 158°F) |
|---|---|
| Operating temperature | -10°C to +40°C (14°F to 104°F) |
| Humidity | 40% to 90% RH (non-condensing) |
| Maximum branch circuit current | 16 A |
| Circuit breaker short-circuit current | Exceeding 35 A |
| Ground connection wire | 18 AWG or wider |
Users are prohibited from installing, calibrating, setting up, or servicing the product. Only qualified and authorized service personnel should perform those tasks.
The operating temperature range is -10°C to +40°C (14°F to 104°F).
The supported humidity range is 40% to 90% RH, non-condensing.
Vishay Transducers warrants instruments to be free from defects in materials and factory workmanship for one year after the date of shipment.