UVP EL Series Ultraviolet Hand Lamps User's Guide

User guide for UVP EL Series ultraviolet hand lamps covering operation, safety, specifications, tube replacement, cleaning, and warranty.


UVP EL Series Ultraviolet Hand Lamps User's Guide - cover page
Brand
UVP
Category
Other Electronics
Document type
User Manual
Language
English
Pages
4
File format
PDF
File size
152 KB
Published
23 November, 2026
Updated
23 November, 2026
MD5 checksum
0CA23BE80CBC1F0FABBDB48C5E40C484

About this manual

The EL Series Ultraviolet Hand Lamps from UVP provide a uniform and intense source of ultraviolet light in four-watt, six-watt, and eight-watt configurations, emitting one of three UV wavelengths or white light for exposure or illumination of materials. All EL Series lamp models can be used with the universal J138 Lamp Stand; four-watt models can be used with the C-10E Cabinet and eight-watt lamps with the C-65 Cabinet.

The guide describes safe operation, including protection from shortwave and midrange UV radiation, power connection, rocker switch use, tube configurations, physical dimensions and electrical ratings, replacement tubes and switches, accessories, tube replacement steps, cleaning, technical support, and warranty terms.

What's inside

  • Introduction
  • Important Safety Information
  • Operation
  • Specifications
  • Replacement Tubes and Switches
  • Accessories
  • Changing the Replacement Tubes
  • Cleaning the EL Series Lamps
  • Maintenance/Repair/Technical Assistance
  • Warranty

Specifications

Lamp power configurationsFour-watt, six-watt, and eight-watt
UV wavelengths254 nm, 302 nm, and 365 nm, plus white light
Four-watt lamp dimensions9.8 in L x 2.5 in H x 3.8 in W (249 x 64 x 97 mm)
Six-watt lamp dimensions11.9 in L x 2.5 in H x 3.8 in W (302 x 64 x 97 mm)
Eight-watt lamp dimensions14.9 in L x 2.5 in H x 3.8 in W (376 x 64 x 97 mm)
Tube configurationTwo UV tubes, or one UV and one white light tube; UVLSM-38 has three tubes
Nominal electrical rating115/60/0.16, 230/50/0.16, or 100/50-60/0.16 (varies by model)
Compatible standUniversal J138 Lamp Stand
Compatible cabinetsC-10E for four-watt models; C-65 for eight-watt lamps

Frequently asked questions

How do I operate the lamp?

Plug the power cord into a properly grounded electrical outlet, then use the rocker switch on top to turn the lamp on or off; multi-wavelength models use the switch to select a wavelength or shut the unit off.

What safety precautions are required?

Shortwave and midrange UV radiation can damage unprotected eyes and skin, so personnel should be properly protected and UV blocking eyewear should be worn.

How do I replace a tube?

Disconnect the lamp from power, remove the end cap screws, slide the chassis and filter out, disconnect the Molex connector, twist the tube a quarter turn to remove and insert the new tube, then reassemble.

What are the lamp dimensions?

Four-watt lamps measure 9.8 by 2.5 by 3.8 inches, six-watt lamps 11.9 by 2.5 by 3.8 inches, and eight-watt lamps 14.9 by 2.5 by 3.8 inches.