User guide for UVP EL Series ultraviolet hand lamps covering operation, safety, specifications, tube replacement, cleaning, and warranty.
The EL Series Ultraviolet Hand Lamps from UVP provide a uniform and intense source of ultraviolet light in four-watt, six-watt, and eight-watt configurations, emitting one of three UV wavelengths or white light for exposure or illumination of materials. All EL Series lamp models can be used with the universal J138 Lamp Stand; four-watt models can be used with the C-10E Cabinet and eight-watt lamps with the C-65 Cabinet.
The guide describes safe operation, including protection from shortwave and midrange UV radiation, power connection, rocker switch use, tube configurations, physical dimensions and electrical ratings, replacement tubes and switches, accessories, tube replacement steps, cleaning, technical support, and warranty terms.
| Lamp power configurations | Four-watt, six-watt, and eight-watt |
|---|---|
| UV wavelengths | 254 nm, 302 nm, and 365 nm, plus white light |
| Four-watt lamp dimensions | 9.8 in L x 2.5 in H x 3.8 in W (249 x 64 x 97 mm) |
| Six-watt lamp dimensions | 11.9 in L x 2.5 in H x 3.8 in W (302 x 64 x 97 mm) |
| Eight-watt lamp dimensions | 14.9 in L x 2.5 in H x 3.8 in W (376 x 64 x 97 mm) |
| Tube configuration | Two UV tubes, or one UV and one white light tube; UVLSM-38 has three tubes |
| Nominal electrical rating | 115/60/0.16, 230/50/0.16, or 100/50-60/0.16 (varies by model) |
| Compatible stand | Universal J138 Lamp Stand |
| Compatible cabinets | C-10E for four-watt models; C-65 for eight-watt lamps |
Plug the power cord into a properly grounded electrical outlet, then use the rocker switch on top to turn the lamp on or off; multi-wavelength models use the switch to select a wavelength or shut the unit off.
Shortwave and midrange UV radiation can damage unprotected eyes and skin, so personnel should be properly protected and UV blocking eyewear should be worn.
Disconnect the lamp from power, remove the end cap screws, slide the chassis and filter out, disconnect the Molex connector, twist the tube a quarter turn to remove and insert the new tube, then reassemble.
Four-watt lamps measure 9.8 by 2.5 by 3.8 inches, six-watt lamps 11.9 by 2.5 by 3.8 inches, and eight-watt lamps 14.9 by 2.5 by 3.8 inches.