Instruction manual for the USCN Life Science C90817Hu chemiluminescent immunoassay kit for quantitative measurement of human Haptoglobin (Hpt).
This instruction manual is for the USCN Life Science C90817Hu chemiluminescent immunoassay kit, designed for in vitro sensitive quantitative measurement of Haptoglobin (Hpt) in human serum, plasma, and other biological fluids. It is for research use only and not for use in diagnostic or therapeutic procedures.
The kit provides 96 tests and includes a pre-coated 96-well strip plate, standards, detection reagents, assay diluents, substrates, wash buffer, and plate sealers. The manual covers reagent and sample handling, storage conditions, standard preparation, and required luminometer settings.
| Tests per kit | 96 |
|---|---|
| Analyte | Haptoglobin (Hpt) |
| Sample types | Human serum, plasma, and other biological fluids |
| Standard stock concentration | 8,000 ng/mL |
| Highest standard | 800 ng/mL |
| Standard dilution series | 800, 400, 200, 100, 50, 25, 12.5, and 0 ng/mL |
| Luminometer lag time | 30.0 secs |
| Luminometer read time | 1.0 sec/well |
| Unopened kit storage | Standard, Detection Reagent A, Detection Reagent B, and 96-well strips plate at -20°C; other reagents at 4°C |
| Opened kit use | Within 1 month if within expiration date |
| Sample storage | 4°C for up to 5 days; otherwise -20°C for up to 1 month or -80°C for up to 2 months |
| Assay Diluent preparation | Dilute 6 mL of 2× concentrate with 6 mL water to make 12 mL |
| Wash Buffer | 30× concentrate |
It is designed for the in vitro sensitive quantitative measurement of Haptoglobin (Hpt) in human serum, plasma, and other biological fluids.
No. It is for in vitro research use only and not for use in diagnostic or therapeutic procedures.
The Standard, Detection Reagent A, Detection Reagent B, and 96-well strips plate should be stored at -20°C upon receipt, while the other reagents should be stored at 4°C.
No. Sample hemolysis will influence the result, so hemolytic specimens cannot be detected.