User manual for ULF01.X.X and ULF03.X.X ultra-low flow sensors, covering safety, design, specs, and installation.
This manual provides instructions for installing and operating ULF ultra-low flow sensors. It includes safety guidelines, technical specifications, connection details for FlowX3 instruments, and information on optional output kits. The sensors are designed for measuring flow of aggressive, solid-free liquids and come in two versions with different flow ranges. Installation requires proper positioning and wiring, and the document specifies compatible materials and environmental limits.
| Flow Rate Range | ULF01 version: 1.5 to 100 l/h (0.0066 to 0.44 gpm); ULF03 version: 6 to 250 l/h (0.0264 to 1.1 gpm) |
|---|---|
| Working Temperature | -10° to 80°C (14° to 176°F) |
| Working Pressure | 5 bar (70 psi) max. @ 22°C (72°F) |
| Viscosity of fluid | 1 to 10 cST |
| Enclosure | IP65 |
| Supply Voltage (Hall version) | 5 to 24 VDC regulated |
| Supply Voltage (Reed version) | none |
| Output Signal | square wave |
| K-factor (ULF01 version, Hall) | 8431 Pulses/Liter (31569 Pulses/U.S. Gallon) |
| K-factor (ULF03 version, Hall) | 3394 Pulses/Liter (12846 Pulses/U.S. Gallon) |
ULF01 version: 1.5 to 100 l/h (0.0066 to 0.44 gpm); ULF03 version: 6 to 250 l/h (0.0264 to 1.1 gpm).
From -10°C to 80°C (14°F to 176°F).
A square wave output proportional to flow velocity.