User manual for the Tianjingsha Pasteurella multocida colorimetric LAMP kit (12-50210): 20 µL reactions at 65°C for 60 min, visual or fluorescent readout.
This is the V1.0 user manual for the Tianjingsha (Beijing Tianjingsha Gene Technology Co., Ltd.) Pasteurella multocida Colorimetric LAMP Kit, catalog number 12-50210. It is a colorimetric loop-mediated isothermal amplification kit for identifying DNA of Pasteurella multocida, a Gram-negative zoonotic pathogen. The kit is supplied with 4×LAMP MasterMix (colorimetric dye method, enzyme to be added), Bst DNA Polymerase 2.0, 20× Pasteurella multocida LAMP primer mix, a non-infectious positive control at 1E4 copies/µL, and ultrapure water, and is enough for 50 reactions of 20 µL.
The manual describes sample DNA preparation with the user's own extraction method, LAMP reaction setup using 14 µL of sample in a 20 µL system, 65°C incubation for 60 minutes in a metal bath, water bath or PCR instrument (with heated lid), and result analysis. Results can be read by eye as a blue color for positive and light blue or colorless for negative, or by fluorescence; gel electrophoresis is strongly discouraged because of aerosol contamination risk. The product is for qualitative analysis only and is for research use only. This manual is written in Chinese.
| Storage and shipping | Shipped at low temperature, stored at -20°C |
|---|---|
| Shelf life | One year |
| Reaction volume | 20 µL |
| Amplification condition | 65°C for 60 minutes |
| Number of reactions | 50 reactions of 20 µL |
| Maximum sample volume | 14 µL per 20 µL reaction |
| Primers | 6 specific primers designed from conserved sequence |
| Positive control concentration | 1E4 copies/µL |
No. The kit uses isothermal amplification and does not require expensive instruments such as a fluorescence PCR machine; a metal bath or water bath can be used, and a fluorescence PCR instrument is optional for real-time detection.
Results are generally obtained within 30 minutes, which the manual states is faster than PCR.
The reaction liquid of the positive controls turns blue, while a negative result stays light blue or colorless; a sample close to the positive control color is positive.