The TestEquity 1007H Temperature/Humidity Chamber operation and service manual provides safety instructions, installation steps, operation guidance…
This manual covers all aspects of using the TestEquity 1007H Temperature/Humidity Chamber, beginning with safety notices and required clearances, followed by detailed installation procedures for power, water, and ventilation. It also explains the chamber’s operation, including front‑panel controls, temperature and humidity set‑points, and the functions of the F4 controller and EZ‑Zone limit controller.
Additional sections describe programming of temperature/humidity profiles, maintenance schedules, troubleshooting, and full technical specifications such as temperature and humidity ranges, power requirements, and physical dimensions. The document also includes wiring diagrams, controller I/O details, and warranty information.
| Temperature Range | –73°C to +175°C |
|---|---|
| Temperature Control Tolerance | ±0.5°C (±0.2°C Typical) |
| Temperature Uniformity | ±0.5°C |
| Humidity Range | 10%RH to 95%RH (10°C to 85°C) |
| Humidity Control Tolerance | ±3% RH |
| Workspace Dimensions | 24" W × 21" H × 24" D (7 cu ft) |
| Weight | 900 lb |
| F4 Controller Digital Inputs | Four (contact closure or DC voltage) |
A minimum clearance of at least 12 inches from both the left and right sides and 24 inches from the rear is required.
The chamber must be configured for either 208 V or 230 V nominal input; the two are not interchangeable.
Humidity is controlled to within ±3 % RH, measured at the control sensor after stabilization.