TestEquity 1007H Temperature/Humidity Chamber Operation and Service Manual

The TestEquity 1007H Temperature/Humidity Chamber operation and service manual provides safety instructions, installation steps, operation guidance…


TestEquity 1007H Temperature/Humidity Chamber Operation and Service Manual - cover page
Brand
TestEquity
Document type
Service Manual
Language
English
Pages
79
File format
PDF
File size
1.1 MB
Published
07 October, 2026
Updated
07 October, 2026
MD5 checksum
751FFAA96AC14364A5DDA0B5B8677AB9

About this manual

This manual covers all aspects of using the TestEquity 1007H Temperature/Humidity Chamber, beginning with safety notices and required clearances, followed by detailed installation procedures for power, water, and ventilation. It also explains the chamber’s operation, including front‑panel controls, temperature and humidity set‑points, and the functions of the F4 controller and EZ‑Zone limit controller.

Additional sections describe programming of temperature/humidity profiles, maintenance schedules, troubleshooting, and full technical specifications such as temperature and humidity ranges, power requirements, and physical dimensions. The document also includes wiring diagrams, controller I/O details, and warranty information.

What's inside

  • Safety Instructions
  • Installation
  • Operation
  • F4 Temperature/Humidity Controller
  • Limit Controller
  • Frequently Asked Questions
  • Specifications
  • Maintenance
  • Theory of Operation
  • Warranty

Specifications

Temperature Range–73°C to +175°C
Temperature Control Tolerance±0.5°C (±0.2°C Typical)
Temperature Uniformity±0.5°C
Humidity Range10%RH to 95%RH (10°C to 85°C)
Humidity Control Tolerance±3% RH
Workspace Dimensions24" W × 21" H × 24" D (7 cu ft)
Weight900 lb
F4 Controller Digital InputsFour (contact closure or DC voltage)

Frequently asked questions

What clearance is required around the chamber for proper ventilation?

A minimum clearance of at least 12 inches from both the left and right sides and 24 inches from the rear is required.

What input voltage configurations are supported by the chamber?

The chamber must be configured for either 208 V or 230 V nominal input; the two are not interchangeable.

What is the humidity control tolerance of the chamber?

Humidity is controlled to within ±3 % RH, measured at the control sensor after stabilization.