This instruction manual provides detailed operation, calibration, troubleshooting, and maintenance procedures for the Teledyne API Model 108E/A Total…
The Teledyne API Model 108E (or M108A) Total Sulfides in CO2 Analyzer is designed to measure mixed sulfur impurities, collectively referred to as Total Sulfides (TS), in carbon dioxide gas. The instrument consists of a modified M100E (M100A) UV Fluorescence SO2 Analyzer with special software and an M501TS high temperature quartz thermal converter.
| Maximum Flow Rate | 1000 cc/min |
|---|---|
| Nominal Flow Rate (CO2) | 625 cc/min |
| Operating Converter Temperature | 1000 °C |
| Maximum Converter Temperature | 1050 °C |
| Power Requirements | 100-120/220-240 VAC 50/60 Hz, 440 watts |
| Weight | 24 (11) lbs. (kg) |
Teledyne API warrants each Product to be free from defects in material and workmanship under normal use and service for a period of one year from the date of delivery.
Sulfur compounds are heated to approximately 1000 °C and converted to SO2 via the reaction: TS + O2 --> SO2.