Installation instructions for the Takara Belmont 0197 dental unit and chair, covering dimensions, specifications, and wiring diagrams.
This guide provides step‑by‑step instructions for installing the Takara Belmont 0197 dental unit and chair. It includes pre‑installation requirements, unpacking procedures, and detailed installation steps for the chair base, J‑Box assembly, stop valves, power supply, and utility sections.
The document also contains hydraulic and electrical diagrams for the chair, unit, and dental light, as well as specifications for power consumption, air and water pressures, weight limits, and environmental operating conditions.
| Power consumption (115 V) | 7.7 A / 6.3 A |
|---|---|
| Power consumption (220 V) | 3.5 A / 3.5 A |
| Air pressure | 0.45 – 0.5 MPa |
| Water pressure | 0.1 – 0.2 MPa |
| Maximum load | 135 kg |
| Protection level against water | IPX1 (IEC60529) |
| Electrical shock protection | Class I Type B |
The unit requires compressed air pressure between 0.45 and 0.5 MPa.
The chair can support a maximum load of 135 kg.
At 115 V the unit consumes 7.7 A (phase) and 6.3 A (neutral).
The unit can operate between +10 °C and +40 °C.