TA Instruments BioDynamic 5100 Series Site Preparation Guide

Site preparation guide for the TA Instruments BioDynamic 5100 Series, covering ideal setup, system components, instrument measurements, and utility…


TA Instruments BioDynamic 5100 Series Site Preparation Guide - cover page
Brand
TA Instruments
Document type
Installation Guide
Language
English
Pages
10
File format
PDF
File size
571 KB
Published
12 October, 2026
Updated
12 October, 2026
MD5 checksum
90729C117960AD57597376AFBB8DEB7C

About this manual

This document serves as a site preparation guide for the ElectroForce BioDynamic 5100 Series Revision A. It provides instructions on selecting an appropriate location with adequate space and a rigid laboratory bench in a vibration-free environment. The guide details ideal placement dimensions, including minimum distance from walls and required table width and depth.

What's inside

  • Ideal Setup
  • System Components
  • Instrument Measurements
  • Power Supply
  • Utility Requirements
  • Site Preparation Checklist

Specifications

Operating temperature18°C (64°F) to 30°C (86°F)
Flow Loop temperature range18°C (64°F) to 40°C (104°F)
Relative humidity requirement40–65% (non-condensing)

Frequently asked questions

What are the power requirements for the instrument?

The instrument requires either 104–132V, 50/60 Hz, 15A or 207–250V, 50/60 Hz, 10A.

What is the minimum distance required from the wall for installation?

The instrument must be placed at a minimum distance of 0.15 m (0.5 ft) from the wall.

What are the required table dimensions for installation?

The bench must have a depth of at least 1.2 m (4 ft) and a width of 1.5 m (5 ft).

Download PDF (571 KB)