Stanford Research Systems SR830 DSP Lock-In Amplifier User Manual

Manual for the Stanford Research Systems SR830 DSP lock-in amplifier, covering operation, specifications, programming, testing, and service information.


Stanford Research Systems SR830 DSP Lock-In Amplifier User Manual - cover page
Brand
Stanford Research Systems
Category
Other Electronics
Document type
User Manual
Language
English
Pages
174
File format
PDF
File size
455 KB
Published
15 November, 2026
Updated
15 November, 2026
MD5 checksum
97A0BAA0E2E8A3988092C7231615FF31

About this manual

The SR830 is a DSP lock-in amplifier with signal, reference, and internal oscillator channels, plus displays, outputs, and auxiliary inputs and outputs. The manual covers safety and preparation, specifications, getting started, operating instructions, front and rear panel functions, programming commands, tests, circuitry, and parts lists.

It describes measurement quantities such as X, Y, R, and θ, time constants, dynamic reserve, filters, phase and frequency settings, data storage and transfer, and interface control through IEEE-488 and RS232.

What's inside

  • General Information
  • Getting Started
  • SR830 Basics
  • Operation
  • Front Panel
  • Rear Panel
  • Programming
  • Testing
  • Circuitry
  • Parts Lists

Specifications

Voltage inputsSingle-ended (A) or differential (A-B)
Current input106 or 108 Volts/Amp
Full-scale sensitivity2 nV to 1 V
Input impedance (voltage)10 MΩ + 25 pF, AC or DC coupled
Gain accuracy±1% from 20°C to 30°C
Input noise6 nV/√Hz at 1 kHz (typical)
Signal filters60 (50) Hz and 120 (100) Hz notch filters (Q=4)
CMRR90 dB at 100 Hz (DC coupled)
Dynamic reserveGreater than 100 dB
Harmonic distortion-80 dB
Reference frequency range1 mHz to 102 kHz
Reference inputTTL or Sine
Phase resolution0.01°
Absolute phase error<1°
Relative phase error<0.01°
Orthogonality90° ± 0.001°
Harmonic detectNxf where N<19999 and Nxf<102 kHz
Acquisition time(2 cycles + 5 ms) or 40 ms, whichever is greater
Time constants (reference > 200 Hz)10 µs to 30 s
Time constants (reference < 200 Hz)up to 30000 s
Harmonic rejection-80 dB
Internal oscillator frequency1 mHz to 102 kHz
Frequency accuracy25 ppm + 30 µHz
Frequency resolution4 1/2 digits or 0.1 mHz
Distortionbelow -80 dBc below 10 kHz
Output impedance50 Ω
Oscillator amplitude4 mVrms to 5 Vrms
Amplitude accuracy1%
Amplitude stability50 ppm/°C
Channel display4 1/2 digit LED display with 40 segment LED bar graph
Offsetup to ±105% of full scale
Expandby 10 or 100
Data buffer16k points
Sample rate512 Hz down to 1 point every 16 seconds
Channel output voltage±10 V full scale
Channel output current10 mA max
Auxiliary output voltage±10.5 V full scale
Auxiliary input voltage±10.5 V full scale
InterfacesIEEE-488 and RS232
Power40 Watts, 100/120/220/240 VAC, 50/60 Hz
Dimensions17"W x 5.25"H x 19.5"D
Weight30 lbs
WarrantyOne year parts and labor

Frequently asked questions

What is the voltage input impedance?

The voltage input impedance is 10 MΩ + 25 pF, AC or DC coupled.

What interfaces does the SR830 include?

IEEE-488 and RS232 interfaces are standard.

What is the power requirement?

The SR830 uses 40 Watts from 100/120/220/240 VAC at 50/60 Hz.

What is the warranty?

The warranty is one year parts and labor on materials and workmanship.

Download PDF (455 KB)