User manual for Spellman's SXR80HR high voltage X-ray generator, covering safety, operation, diagnostics, and maintenance.
This manual provides operating instructions and safety guidelines for the Spellman SXR80HR high voltage X-ray generator. It includes sections on initial inspection, operation modes, diagnostics, and factory service. The device is a sealed unit with internal HVPS, X-ray tube, and oil encapsulant for insulation and cooling. Safety precautions emphasize grounding, discharge procedures, and qualified servicing. Remote control via RS-232 is supported for voltage, current, and interlock control.
The generator features internal fault protection, LED status indicators, watchdog timer, and oil temperature monitoring. It is designed for OEM applications with up to 80kV at 100W, and includes analog and digital monitoring ports. Maintenance requires qualified personnel due to high voltage risks.
| Maximum Voltage | 80kV |
|---|---|
| Maximum Power | 100W |
| Output Voltage | +390VDC |
| Cooling Method | Convection cooling augmented by external fan |
| Interlock Control | External interlock via J2-11 to J2-12 |
The SXR80HR is designed to power its internal X-ray tube up to 80kV at 100W.
The device uses convection cooling augmented by an external fan provided by the customer.