The YG141 digital fabric thickness gauge manual provides setup, operation, maintenance, and calibration instructions.
About this manual
The YG141 digital fabric thickness gauge is designed for measuring the thickness of various textiles according to GB/T3820-1997. It features an electric lift, lever balance, automatic timing, and a digital LCD display, allowing single or continuous measurements with selectable pressure times of 10 s or 30 s.
The manual covers unboxing, inspection, proper use of the pressure foot and reference plate, zeroing the display, battery replacement, troubleshooting, and periodic calibration to maintain measurement accuracy. This manual is written in Chinese.
What's inside
- Product Overview
- Structure Diagram
- Purpose
- Technical Specifications
- Instrument Structure
- Unboxing and Inspection
- Operation
- Dial Indicator Instructions
- Troubleshooting
- Calibration
Specifications
| Measurement range | 0.02–10 mm |
|---|
| Resolution | 0.001 mm |
|---|
| Accuracy | ±0.01 mm |
|---|
| Foot area options | 100, 2000, 2500, 10000 mm² |
|---|
| Foot diameter options | 11.28, 50.46, 56.42, 112.84 mm |
|---|
| Pressure time | 10 s or 30 s |
|---|
| Foot descent speed | 1.89 mm/s |
|---|
| Instrument weight | ≈25 kg |
|---|
| Dimensions | 450 × 216 × 305 mm |
Frequently asked questions
How do I zero the digital display?
Enter relative measurement mode (display shows INC), then press the ABS key twice to zero the display.
What are the recommended operating temperature and humidity?
Operate the gauge between 0 °C and 40 °C with relative humidity not exceeding 80 %.