User manual for SENSIT SCAL-400 Calibration and Charging Station, which automatically charges, bump tests, and calibrates up to five SENSIT P400 gas monitors.
The SENSIT SCAL-400 Calibration and Charging Station is designed to automatically charge, bump test, and calibrate up to five SENSIT® P400 Gas Monitors simultaneously. It features a touch screen display and RJ45 intranet connection for networked operation. The station includes SCAL-D Data Management Software for tracking test results and compliance reports. It is made in the USA and comes with a 2-year technical support period.
The station operates with Windows 7 and Windows XP, and its software can generate reports, alerts, and store data in SQL Server. It includes a power cord, communication cable, and installation guide. The device is compact, weighing 5.75 lb and measuring 24” x 7” x 3”.
| Size | 24” x 7” x 3” (61cm x 17.8cm x 7.6cm) |
|---|---|
| Weight | 5.75 lb. (2.6 kg) |
| Operational Temperature | 40 – 122 F (5 to 50° C) |
| Max. Inlet Pressure | 20 PSI |
| Power | 100-240 VAC |
It is designed for use with SENSIT® P400 Gas Monitors that detect CH4, CO, O2, H2S, HCN, NO2, and SO2.