Sanxiang 2019-nCoV Nucleic Acid Detection Kit (Fluorescent PCR) Instructions

This manual provides instructions for using the Sanxiang 2019-nCoV nucleic acid detection kit for qualitative diagnosis of SARS-CoV-2 via fluorescent PCR.


Sanxiang 2019-nCoV Nucleic Acid Detection Kit (Fluorescent PCR) Instructions - cover page
Brand
Sanxiang
Category
Other
Document type
User Manual
Language
Chinese
Pages
2
File format
PDF
File size
257 KB
Published
17 October, 2026
Updated
17 October, 2026
MD5 checksum
B56CFB77229769F5C62F595972365145

About this manual

The Sanxiang 2019-nCoV Nucleic Acid Detection Kit (Fluorescent PCR) is designed for the in vitro qualitative detection of SARS-CoV-2 ORF1ab and N genes in pharyngeal swab and alveolar lavage fluid samples. The kit utilizes real-time fluorescent quantitative RT-PCR technology with dual-target gene design and an internal control to monitor sample collection and extraction processes.

The manual details the kit's composition, storage conditions, and specific requirements for sample types and handling. It outlines the step-by-step inspection method, including reagent preparation, sample processing, PCR amplification parameters, and result analysis. Quality control criteria for positive and negative controls are specified to ensure valid experimental results. This manual is written in Chinese.

What's inside

  • Product Name
  • Expected Use
  • Inspection Principle
  • Main Components
  • Storage Conditions
  • Sample Requirements
  • Inspection Method
  • Quality Control
  • Result Interpretation
  • Product Performance Indicators

Specifications

Packaging Specification24 tests/box
Minimum Detection Limit200 copies/mL
Applicable InstrumentSLAN-96P
Sample TypesPharyngeal swab, Alveolar lavage fluid
Intrabatch Precision CV≤5%

Frequently asked questions

What is the minimum detection limit of this kit?

The minimum detection limit of this kit is 200 copies/mL.

What are the quality control requirements for the negative control?

For the negative control, there should be no Ct value in the FAM, ROX, or internal standard (HEX) channels, or the Ct value must be greater than 40.

How is a sample interpreted as positive?

A sample is reported as positive if a typical S-shaped amplification curve is detected in the FAM or ROX channel with a Ct value less than or equal to 40.