The Rockland Scientific MicroRider-1000LP Instrument Manual provides setup, operation, communication, maintenance, and technical specifications for the…
The MicroRider-1000LP is an internally recording microstructure instrument designed for deployment on platforms such as CTD rosettes, AUVs, or gliders. It operates to a maximum depth of 1000 m and is powered by a 12 VDC supply, drawing about 1 W when active.\n\nThe manual covers receipt of goods, software installation, instrument communication, detailed mechanical and electrical overviews, sensor configurations, operation procedures, maintenance disassembly, storage preparation, and an appendix with external connection details.
| Maximum operating depth | 1000 m |
|---|---|
| Power consumption (on) | ≈ 1 W |
| Power consumption (off) | 0 W |
| Operating voltage range | 9.5 V – 17 V |
| Typical current draw | 100 mA at 12 V |
| Data storage medium | Compact Flash card |
| USB transfer speed | ≈ 200 kB/s |
The instrument is rated for a maximum depth of 1000 meters.
It runs from any 12 VDC supply, drawing about 1 W when on and 0 W when off, with an operating voltage range of 9.5 V to 17 V.
There are two methods: a logic‑level signal (0 V turns it on, 2.4‑5 V turns it off) or a current of 2‑10 mA through an optical coupler for ON and <1 mA for OFF.