RIGOL RP1003C/RP1004C/RP1005C Current Probe User Manual

This manual provides operating instructions, safety precautions, and technical specifications for the RIGOL RP1003C, RP1004C, and RP1005C current probes.


RIGOL RP1003C/RP1004C/RP1005C Current Probe User Manual - cover page
Brand
Rigol
Category
Other Electronics
Document type
User Manual
Language
Chinese
Pages
59
File format
PDF
File size
1.3 MB
Published
16 October, 2026
Updated
16 October, 2026
MD5 checksum
B90BFCE6F35B9302CE833BA5FAA2C7ED

About this manual

This user manual covers the RIGOL RP1003C, RP1004C, and RP1005C current probes, detailing their components, usage methods, and safety precautions. It includes step-by-step instructions for measurement preparation, demagnetization, and zero adjustment, along with important notes on handling the sensor head and cables to ensure accuracy and prevent damage. This manual is written in Chinese.

What's inside

  • Safety Precautions
  • Current Probe Introduction
  • Component Overview
  • Measurement Preparation
  • Demagnetization and Zero Adjustment
  • Measurement Steps
  • Specifications
  • Appendix: Frequency Response
  • Appendix: Input Impedance

Specifications

Bandwidth (RP1003C)DC to 50MHz (-3dB)
Bandwidth (RP1004C)DC to 100MHz (-3dB)
Rise Time (RP1003C)≤7ns
Maximum Continuous Input Range30Arms
Maximum Peak Current Value50A peak (non-continuous)
Gain0.1V/A
Amplitude Accuracy±1.0%rdg±1mV (≤30Arms)
Operating Temperature0°C to 40°C

Frequently asked questions

What is the maximum continuous input range for the RP1003C and RP1004C?

The maximum continuous input range is 30Arms, with reference to Appendix 2 for the relationship between max input current and frequency.

What is the maximum continuous input current for the RP1003C and RP1004C models?

The maximum continuous input range is 30Arms.

What is the maximum peak current value for these current probes?

The maximum peak current value is 50A peak for non-continuous use.