Quick guide for the RIGOL DS4000E series digital oscilloscope covering safety, first-time setup, probe connection, probe compensation and remote control.
This Quick Guide from RIGOL introduces the basic information needed when using the DS4000E series digital oscilloscope for the first time. It covers the guaranty and declaration, general safety summary, safety notices and symbols, care and cleaning, and a document overview with format conventions.
The guide describes general inspection of the packaging, instrument and accessories, gives an overview of the front panel, rear panel and user interface, and explains how to prepare the oscilloscope for use: removing the front panel cover, connecting the power supply, power-on inspection, connecting the probe, function inspection for the passive probe, probe compensation, using the built-in help system and remote control. It also points to the RIGOL website (www.rigol.com) for more product information and the latest manuals.
| DS4024E analog bandwidth | 200 MHz |
|---|---|
| DS4024E analog channels | 4 |
| DS4014E analog bandwidth | 100 MHz |
| DS4014E analog channels | 4 |
| AC power input | 100 V-127 V, 45 Hz-440 Hz or 100 V-240 V, 45 Hz-65 Hz |
Restore factory settings with the Default key, connect the probe between CH1 and the probe compensation signal output terminal, press AUTO, then adjust the low-frequency compensation adjustment hole on the probe with a non-metallic screwdriver until the displayed square wave is correctly compensated.
Restore the oscilloscope to factory settings, connect the probe's ground alligator clip to the ground terminal and the probe tip to the probe compensation signal output terminal on CH1, then press AUTO and observe that a square wave is displayed.
Press the Default key and then the OK/confirm key to restore the oscilloscope to factory settings.
No, the probe compensation signal is only for probe compensation adjustment and must not be used for calibration.