RIGOL DM3068 digital multimeter quick guide covering safety requirements, front and rear panels, measurement connections, rack mounting, and troubleshooting.
The RIGOL DM3068 Digital Multimeter Quick Guide covers safety requirements and a general safety summary, safety terms and symbols, care and cleaning, and environmental considerations. It also describes general inspection, handle adjustment, appearance and dimensions, the front and rear panels, the user interface, first use of the multimeter, and measurement connections.
The guide explains use of the built-in help system and a rackmount kit, including installation steps with M4 and M6 screws for a standard 19-inch cabinet. It lists troubleshooting and states input terminal protection limits and IEC II overvoltage protection conditions.
| Maximum measurable voltage (HI-LO) | 1000 VDC or 750 VAC |
|---|---|
| LO-ground protection limit | 500 Vpk (relative) |
| HI Sense-LO Sense/200 mA protection limit | 200 Vpk |
| LO Sense/200 mA-LO protection limit | 0.5 Vpk |
| 10 A terminal maximum current | 10 A (internal fuse) |
| LO Sense/200 mA terminal maximum current | 250 mA (internal fuse) |
| Measurement Category I maximum voltage | 1000 Vpk |
| Measurement Category II mains voltage | up to 300 VAC |
| Rack height | 3U |
| Rack cabinet standard | 19-inch machine cabinet |
The HI-LO protection limit is 1000 VDC or 750 VAC, which is also the maximum measurable voltage.
No. The guide says not to connect the 10 A and LO Sense/200 mA input terminals into the current measuring circuit at the same time.
Disconnect the instrument from all power sources first.
The front panel is fixed using four M4 screws.