This manual provides operating procedures, component storage conditions, and data analysis methods for the Proteintech Double Antibody Sandwich ELISA Kit.
This document is the operation manual for the Proteintech Double Antibody Sandwich ELISA Kit. It details the components included in the kit, their storage conditions, and the necessary laboratory equipment for the experiment. The guide covers sample preparation for various types including serum, plasma, and cell lysates, as well as reagent preparation steps.
| Microplate Capacity | 8 wells x 12 strips |
|---|---|
| Standard Storage | 4°C for 6 months |
| Detection Antibody Concentration | 100 X |
| Wash Buffer Concentration | 20 X |
| TMB Volume | 12 mL |
| Stop Solution Volume | 12 mL |
All kit components should be stored at 4°C for up to 6 months.
Required equipment includes a microplate reader capable of reading 450 nm and 630 nm, a high-precision pipette, and a plate washer.
The OD values should be fitted using a four-parameter logistic (4 PL) method using software such as Origin or ELISACalc.
The Stop Solution is added to terminate the reaction, changing the color from blue to yellow.