Prokit MT-4618 sound level meter user's manual: battery setup, A/C weighting, FAST/SLOW response, MAX hold, 30-130 dB range and specifications.
The Prokit MT-4618 is a digital sound level meter used to measure the ambient sound level around factories, workshops, schools, residential areas, offices, roads and audio equipment, and for noise engineering, product quality control and health prevention. It complies with IEC651 Type 2 and ANSI S1.4 Type 2, covers a measuring scope of 30 to 130 dB with automatic shifting, and provides A and C weighting networks, FAST/SLOW response rates, a MAX locking function and a backlight.
This first edition (2020) user's manual gives the safety instructions and operating environment, explains the LCD display symbols, lists the functional characteristics and specifications, and then describes preparations before use and the basic usage methods: installing four AAA batteries, selecting a measurement gear level when UNDER or OVER appears, choosing the A or C weighting mode and FAST or SLOW response, using the MAX key and lighting the backlight. Closing notes cover use around hot or humid conditions, battery removal and the anti-wind ball. The document contains English text followed by a traditional Chinese section.
| Compliance | IEC651 Type 2 and ANSI S1.4 Type 2 |
|---|---|
| Measuring range | 30 to 130 dBA, 30 to 130 dBC |
| Sound pressure accuracy | ±1.5 dB at 94 dB @ 1 kHz; ±5 dB at 94 dB @ 8 kHz |
| Frequency response | 30 Hz - 8 kHz |
| Dynamic range | 50 dB for each measurement gear level |
| Measurement gear levels | 30-80, 40-90, 50-100, 60-110, 70-120, 80-130 dB (6 levels) |
| Frequency weighting | A and C weighting networks |
| Dynamic characteristic | FAST 125 ms, SLOW 1 sec |
| Microphone | Polarized capacitive microphone |
| Digital display | 4-digit, resolution 0.1 dB, sampling rate 2 times/sec |
| Power supply | AAA x 4 batteries of 1.5 V each |
| Operating temperature | 0 to +40 °C |
| Storage temperature | -10 to +60 °C |
| Dimensions | 193 (L) x 60 (W) x 29 (H) mm |
| Accessories | Instruction manual, cotton ball |
It measures from 30 to 130 dB, with automatic shifting that selects the best gear level within that range.
Four AAA 1.5 V alkaline batteries, installed in the battery holder after opening the back cover with a Phillips head screwdriver.
UNDER means the gear level of the instrument is too high, and OVER means it is too low, so the reading is not accurate until the level is adjusted.
Select the A weighting mode, which simulates the hearing characteristics of human ears; select C weighting to measure the actual sound level.
The manual states it complies with IEC651 Type 2 and ANSI S1.4 Type 2.