Ohio Semitronics CTR Series True RMS Current Transducer Calibration Manual

Calibration guide for Ohio Semitronics CTR series true RMS current transducers, covering setup, zero adjustment, calibration, and linearity checks.


Ohio Semitronics CTR Series True RMS Current Transducer Calibration Manual - cover page
Brand
OHIO SEMITRONICS
Document type
Service Manual
Language
Chinese
Pages
20
File format
PDF
File size
267 KB
Published
08 November, 2026
Updated
08 November, 2026
MD5 checksum
1C64086F9CC64943E9FF908D481F6E1E

About this manual

This manual provides step-by-step instructions for calibrating Ohio Semitronics CTR series true RMS current transducers. It includes details on required test equipment, connection diagrams, zero and calibration adjustments, and linearity verification procedures. The document also specifies model-specific output options and accuracy tolerances.

Calibration is performed using an AC sine wave current source and precision measuring instruments. Adjustments are made via trimpots accessible through the lid. The manual includes tables for verifying linearity and lists recommended calibration equipment from Ohio Semitronics and third-party vendors. This manual is written in Chinese.

What's inside

  • Description
  • Specifications
  • Calibration
  • Test Equipment Required
  • Connections and Adjustments
  • Zero Adjust
  • Calibration Adjust
  • Linearity Check
  • Calibration Equipment Recommended
  • Appendix A Harmonic content

Specifications

Frequency Range48 to 420 Hertz
Accuracy±0.25% of full scale (except CTR-005*Y34 and CTR-010*Y34: ±0.5% @ 60Hz)
Output OptionsOption B: 0 to 1mAdc; Option D: 0 to 10Vdc; Option E: 4 to 20mAdc; Option X5: 0 to 5Vdc
Instrument Power120 volts AC (standard); 220 volts AC (option -22)

Frequently asked questions

What is the operating frequency range for the CTR Series transducers?

The transducers measure true RMS values over an operating range of 50 to 400 hertz.

How long must the transducer warm up before making adjustments?

You must apply instrument power and allow the transducer to warm up for 15 minutes.

What is the recommended frequency for calibration?

You should calibrate the transducer at the frequency at which it will be used.