Nikon Ti2 Eclipse microscope user manual covering startup, software, cameras, objectives, imaging acquisition, autofocus, data saving, and TIRF alignment.
This Nikon Ti2 Eclipse user manual describes how to turn the system on and off, start NIS -ELEMENTS, select a camera, view through the microscope, and set up live cell imaging. It also lists the available objectives and their NA, working distance, contrast method, and application.
The guide covers phase contrast acquisition, Köhler adjustment, automated ND acquisitions, multichannel and fluorescence filter setup, large image and timelapse acquisition, PFS autofocus, multipoint and Z-acquisition, scale bars, data saving, and TIRF alignment.
| TIS camera resolution | 2048 x 2048 pixels |
|---|---|
| Fi3 camera resolution | 2880 x 2048 pixels |
| 10x phase contrast objective | Air Plan Fluor; NA 0,30; WD 15,2 mm; Ph 1 / phase contrast |
| 20x phase contrast objective | Air Plan Fluor (coverslip correction); NA 0,45; WD 7,4 mm; Ph 1 / phase contrast |
| 20x fluorescence objective | Air Plan Apochromat lambda; NA 1,20; WD 1,0 mm; fluorescence |
| 40x fluorescence objective | Air Plan Apochromat lambda (coverslip correction); NA 0,95; WD 0,21 mm; fluorescence |
| 60x TIRF / fluorescence objective | Oil CFI Apo TIRF (coverslip and temperature correction); NA 1,49; WD 0,12 mm; TIRF / fluorescence |
| 100x fluorescence objective | Oil CFI Plan Apochromat lambda; NA 1,45; WD 0,13 mm; fluorescence |
Use the TIS camera for imaging of fluorescent samples.
Use the Fi3 camera for brightfield, phase contrast and histological samples.
The 60x and 100x objectives must be used with immersion oil; the other objectives are dry.
No password is needed for the user account.