User manual for the Microlife MT1P21-PRO digital thermometer: oral and underarm use, °F/°C switching, cleaning, battery change and specifications.
The Microlife MT1P21-PRO is a predictive clinical digital electronic thermometer for the intermittent measurement and monitoring of adult and pediatric body temperature, taken orally or under the arm (axillary). It can be used with or without a single-use tip cover and is not intended for neonatal use. The document includes general warnings, such as contacting a physician for prolonged fever and keeping batteries away from children.
The manual explains how to take a temperature, including the function check showing "188.8" and the last recorded measurement, oral readings in about 30 seconds and underarm readings in about 5 minutes. It also covers the beep signals for readings at or below 99.5 °F (37.5 °C) and above, switching between °F and °C, cleaning with 70% isopropyl alcohol or on the top rack of a dishwasher, replacing the 1.5V LR41 battery, the 5-year limited warranty, technical specifications and a test protocol.
| Measurement range | 90.0 °F to 109.9 °F (32.0 °C to 43.9 °C) |
|---|---|
| Measurement accuracy | ±0.2 °F (±0.1 °C) between 93.2 °F (34 °C) to 107.6 °F (42 °C) |
| Display | Liquid crystal display (LCD) with four digits |
| Smallest unit of display | 0.1 °F (0.1 °C) |
| Auto off | With reading: after 9 minutes of inactivity; without reading: after 1 minute of inactivity |
| Operating temperature | 50 °F (10 °C) to 104 °F (40 °C), humidity 15% to 95% (non-condensing) |
| Storage temperature | -13 °F (-25 °C) to 140 °F (60 °C), humidity 15% to 95% (non-condensing) |
| Battery | 1.5V, LR41 (or equivalent) |
| Ingress protection | IP67: totally protected against dust ingress; protected against the effect of immersion between 15 cm and 1 m |
| Expected product life | 5 years |
| Warranty | 5 years by Microlife USA Inc against manufacturer defects for the original purchaser; batteries and case not covered |
About 30 seconds with proper placement under the tongue; an open mouth or improper placement could make the reading take longer.
With the thermometer turned off, press and hold the power button for 5 seconds, then press the ON/OFF button to toggle to the opposite temperature setting.
Yes, place the thermometer on the top rack only and run the dishwasher on a normal cleaning cycle; high heat drying is not recommended.
A 1.5V LR41 battery or equivalent, with the plus (+) sign facing upward.