This document introduces the Merck Millipore Elix Reference pure water system, detailing its design elements, technical features, water quality monitoring…
The Merck Millipore Elix Reference pure water system is designed to produce stable and reliable II-grade pure water using municipal tap water as input. It combines patented Elix continuous current deionization technology with advanced purification methods, including Progard pre-treatment and smart reverse osmosis (RO), to ensure consistent water quality without the need for chemical regeneration or resin replacement.
The system offers low operating costs and predictable performance through its energy-efficient design and reduced water consumption. It features advanced water quality monitoring with high-precision sensors and optional Millitrack software for data management, remote access, and long-term archiving, ensuring compliance with strict laboratory standards like ISO 3696 and USP. This manual is written in Chinese.
| Resistivity (Typical) | >5 MΩ·cm (at 25°C), typical value 10-15 MΩ·cm |
|---|---|
| Total Organic Carbon (TOC) | <30 ppb |
| Organic Removal Efficiency | 95-99% (MW > 200 Dalton) |
| Waste Water Recovery Rate | Up to 50% |
The system produces pure water that meets or exceeds ISO 3696 (Grade 2), GB/T 6682-2008, ASTM D1193, and relevant pharmacopoeia regulations.
No, it uses patented continuous current deionization technology that eliminates the need for adding softeners, chemical regeneration, or changing DI columns.
The Elix module consumes only as much electricity as a standard incandescent light bulb.