Linchy LH-600C Hydrogen Generator User Manual

Linchy LH-600C hydrogen generator user manual covering specifications, installation, operation, maintenance and troubleshooting.


Linchy LH-600C Hydrogen Generator User Manual - cover page
Brand
Linchy
Category
Other
Document type
User Manual
Language
Chinese
Pages
12
File format
PDF
File size
503 KB
Published
16 November, 2026
Updated
16 November, 2026
MD5 checksum
5885C6505402FA1B3DD6177EE8333165

About this manual

The LH-600C hydrogen generator manual describes an SPE electrolysis-based laboratory hydrogen source that produces high-purity hydrogen from deionized water. It covers the instrument's parts, technical specifications, main components, operating requirements, acceptance and installation, common fault repair, after-sales service, and packing list.

The manual specifies an output flow of 0-600 ml/min, output pressure of 0.3 MPa, hydrogen purity of 99.999%, and input power of 150 W. It also requires deionized or secondary distilled water with resistivity above 1 MΩ·cm, water-tank filling above two-thirds of its 3.2 L volume, and use in a ventilated, non-closed environment. This manual is written in Chinese.

What's inside

  • General description
  • Working principle and process flow
  • Instrument parts and panel layout
  • Technical specifications
  • Main components
  • Operating requirements
  • Acceptance, installation and use
  • Common fault repair
  • After-sales service
  • Packing list

Specifications

Output flow0-600 ml/min
Output pressure0.3 MPa
Hydrogen purity99.999%
Overpressure protection0.4 MPa
Power supply220 V ±15%, 50-60 Hz
Input power150 W
Dimensions (L×W×H)360×235×380 mm
Weight<14 kg
Water tank volume3.2 L
Water resistivity>1 MΩ·cm
Minimum operating pressure0.012 MPa
Drying systemTwo-stage: color-changing silica gel and molecular sieve

Frequently asked questions

Can the LH-600C be used in a closed room?

No; the manual states that the hydrogen generator must not be used in a closed room.

What water is required for electrolysis?

Use soft water, either deionized or secondary distilled water, with resistivity greater than 1 MΩ·cm.

What is the minimum operating pressure?

The minimum operating pressure is 0.012 MPa.

What must be replaced together with the color-changing silica gel?

When replacing the first-stage color-changing silica gel, the second-stage molecular sieve must also be replaced.