Linchy LH-600C hydrogen generator user manual covering specifications, installation, operation, maintenance and troubleshooting.
The LH-600C hydrogen generator manual describes an SPE electrolysis-based laboratory hydrogen source that produces high-purity hydrogen from deionized water. It covers the instrument's parts, technical specifications, main components, operating requirements, acceptance and installation, common fault repair, after-sales service, and packing list.
The manual specifies an output flow of 0-600 ml/min, output pressure of 0.3 MPa, hydrogen purity of 99.999%, and input power of 150 W. It also requires deionized or secondary distilled water with resistivity above 1 MΩ·cm, water-tank filling above two-thirds of its 3.2 L volume, and use in a ventilated, non-closed environment. This manual is written in Chinese.
| Output flow | 0-600 ml/min |
|---|---|
| Output pressure | 0.3 MPa |
| Hydrogen purity | 99.999% |
| Overpressure protection | 0.4 MPa |
| Power supply | 220 V ±15%, 50-60 Hz |
| Input power | 150 W |
| Dimensions (L×W×H) | 360×235×380 mm |
| Weight | <14 kg |
| Water tank volume | 3.2 L |
| Water resistivity | >1 MΩ·cm |
| Minimum operating pressure | 0.012 MPa |
| Drying system | Two-stage: color-changing silica gel and molecular sieve |
No; the manual states that the hydrogen generator must not be used in a closed room.
Use soft water, either deionized or secondary distilled water, with resistivity greater than 1 MΩ·cm.
The minimum operating pressure is 0.012 MPa.
When replacing the first-stage color-changing silica gel, the second-stage molecular sieve must also be replaced.