LIGO manual for Diagnostic Breadboard computer control: field box cabling and channels, real-time code model, configuration, calibration and measurements.
This manual describes how to perform diagnostic laser beam measurements with the computer control of the Diagnostic Breadboard (DBB). It covers the field box that connects the DBB with the A/D and D/A cards, the Simulink model used for real time code generation, and the MEDM screens, and it gives instructions for calibrating the field box and the A/D and D/A cards and for performing the laser beam measurements.
The document starts with definitions and acronyms and then describes the field box, its cabling, channel assignment and whitening/de-whitening filters. Later sections cover the real time code model and its subsystems, the control screen, the various configuration items (scanning, TPD and RPD photodiodes, prealignment, demodulation phase, PMC length control, auto alignment and interlock), calibration of photodiodes, PZT and error signals, and measurements of relative power noise, frequency noise, pointing noise, beam quality and data acquisition.
| A/D and D/A channels required for control | 24 A/D and 16 D/A channels |
|---|---|
| Minimum A/D card channels | At least 24 |
| Minimum D/A card channels | At least 16 |
| Auxiliary inputs | 3 inputs, approx. -10 V to 10 V |
| Supported multiplexing shutters | Two shutters (Thorlabs SH05) |
It is an electronic module placed in the electronic crate of the DBB and is the connection between the electronic signals of the DBB and the computer control system.
24 A/D channels and 16 D/A channels are necessary.
The switches have to be in the 'ADD' position when the two control modules are in the crate.
They must never be operated at the same time, because the modules will try to drive the same lines on the crate bus.