LeiCi WZS-185A Turbidimeter User Manual

This manual covers the installation, operation, maintenance, troubleshooting, and standard solution preparation for the LeiCi WZS-185A Turbidimeter.


LeiCi WZS-185A Turbidimeter User Manual - cover page
Brand
LeiCi
Document type
User Manual
Language
Chinese
Pages
18
File format
PDF
File size
469 KB
Published
26 October, 2026
Updated
26 October, 2026
MD5 checksum
FB1CB4C595DB61937FE8F7E93401BD2A

About this manual

The document serves as a user guide for the WZS-185A Turbidimeter manufactured by Shanghai Instrumentation Science Instrument Co., Ltd. It details the unboxing process, instrument structure, and installation procedures including USB communication setup.

The manual provides comprehensive instructions on operating the device, utilizing its functional menu for settings like clock adjustment and data management, and performing measurements based on ISO 7027 standards. It also includes sections on maintenance, common fault troubleshooting, and appendices for driver installation and standard solution preparation. This manual is written in Chinese.

What's inside

  • Instrument Installation
  • Operation Overview
  • Functional Menu
  • Data Management
  • Maintenance
  • Troubleshooting
  • USB Driver Installation
  • Standard Solution Preparation

Specifications

Measurement PrincipleScattering-Transmittance Light Measurement
Standard ComplianceISO 7027
Measurement Range(0~20.00)NTU, (20.0~200.0)NTU, (200~1000) NTU
Zero DriftNot exceeding ±0.5%FS/30min
RepeatabilityNot greater than 0.5%
Show Value ErrorNot exceeding ±6%
Operating Temperature(5~35)℃
Dimensions250×95×60mm (L×W×H)
WeightApprox. 0.4kg

Frequently asked questions

What is the measurement principle of the WZS-185A?

The instrument uses a scattering-transmittance light measurement principle with automatic chromaticity compensation.

How many groups of measurement data can be stored?

The instrument can store up to 2000 sets of measurement data.

What is the communication format for USB connection?

The standard RS-232 communication format is 9600bps, no parity check, 8 data bits, and 1 stop bit.

What causes large errors in measurement results?

Common causes include scratches on the turbidity bottle, attached substances, fingerprints, incorrect standard solution calibration, or improper insertion position.