Langer EMV P331 L-ESD ESD Generator User Manual

User manual for the Langer EMV P331 L-ESD ESD generator probe, covering design, waveform parameters, system setup, safety, and specifications.


Langer EMV P331 L-ESD ESD Generator User Manual - cover page
Brand
Langer EMV
Category
Test Set
Document type
User Manual
Language
English
Pages
10
File format
PDF
File size
1.1 MB
Published
17 November, 2026
Updated
17 November, 2026
MD5 checksum
55AFB235FDA42AD9DED54DD9D7C83FBF

About this manual

The Langer EMV P331 L-ESD is an ESD generator probe used to inject ESD pulses into IC pins through conductors, either directly or through coupling networks for interfaces such as USB, LVDS, and Ethernet. It is operated with the BPS 203 burst power station, which supplies high voltage and controls the probe.

The manual covers the probe design and function, waveform characteristics and verification, system setup with the ICE1 test bed, safety instructions, warranty, technical specifications, and the scope of delivery.

What's inside

  • P331 L-ESD generator
  • Design and function of the P331 L-ESD probe
  • Characteristics
  • System set-up
  • Verifying the waveform
  • Safety instructions
  • Warranty
  • Technical specifications
  • Scope of delivery

Specifications

Pulse risetime200 ps
Pulse width5 ns
Voltage range100 V - 6 kV
Frequency range0.1 Hz – 10 Hz
Dimensions (width/height/depth)79/41/40 mm
Weight0.25 kg
Energy storage capacity100 pF
Internal resistance100 Ω
Current per kilovolt10.6 A/kV
Minimum operating voltage200 V
Required power stationBPS 203
Calibration intervalevery two years

Frequently asked questions

What is the Langer EMV P331 L-ESD used for?

It generates ESD pulses for injection into the device under test via conductors, coupling ESD into IC pins directly or indirectly through coupling networks.

Which power station is required to operate the P331 L-ESD probe?

The probe can only be operated together with the BPS 203 burst power station, which generates the high voltage and controls the probe.

How often must the probe be calibrated?

If the waveform does not deviate from the given parameters, the probe only has to be calibrated every two years by Langer EMV-Technik GmbH.

How is the ESD pulse waveform verified?

The SM 02-01 shunt is inserted into the GNDA 02 ground adapter and its SMA output is connected to the 50 Ω input of an oscilloscope with a 26 dB attenuator set to x20.