KKPT PT2DL Series Dual-Channel Signal Conditioner User Manual

KKPT PT2DL series dual-channel signal conditioner user manual: panel menus, parameter settings, specifications, wiring and MODBUS-RTU registers.


KKPT PT2DL Series Dual-Channel Signal Conditioner User Manual - cover page
Brand
KKPT
Category
Converter
Document type
User Manual
Language
Chinese
Pages
2
File format
PDF
File size
926 KB
Published
16 November, 2026
Updated
16 November, 2026
MD5 checksum
529F5E38A13B40C6BA8D3B7A702010D3

About this manual

The KKPT PT2DL series is a dual-channel signal conditioner that displays two input values at the same time on an LCD with fault status indication. It accepts RTD, thermocouple, voltage, current, millivolt and remote pressure gauge signals, provides two mutually isolated output channels, and communicates over MODBUS-RTU at rates up to 57600 bps.

The manual describes the front panel keys and menu operation, including locking all menus with a password, then covers alarm, function, scale, communication and calibration parameter settings. It also gives the technical specifications and input signal table, explains the model code, and includes sections on wiring, maintenance and the 18-month quality warranty, plus appendices on remote pressure gauge scaling, D/A output correction and MODBUS register definitions. This manual is written in Chinese.

What's inside

  • Instrument Panel Description
  • Menu Operation
  • Parameter Setting
  • Alarm Setting
  • Function Setting
  • Scale Setting
  • Communication Parameter Setting
  • Technical Specifications
  • Wiring
  • Maintenance and Warranty

Specifications

Input signal typesRTD, thermocouple, voltage, current, millivolt and remote pressure gauge signals
Thermocouple input typesK, J, E, T, R, B, S, N, U (WRe3-WRe25)
RTD input typesPt100, Cu50, G53, B1, B2
CommunicationMODBUS-RTU, up to 57600 bps
Basic accuracyOther signals ±0.2%; RTD ±0.2% for ranges ≥10 Ω and ±0.5% below; thermocouple ±0.2% for ranges ≥3 mV and ±0.5% below
Operating temperature-20 °C to 60 °C, humidity <95%
Storage temperature-40 °C to 85 °C
Isolation voltage1000 Vrms
Power supply20 VDC to 35 VDC
Power consumptionApproximately 2 W at 24 VDC
Current output load≤650 Ω
Voltage output load≥2 kΩ
Dimensions22.5 W × 110 H × 115 L mm

Frequently asked questions

What is the factory password of the instrument?

The factory password is 2000, and it unlocks any menu.

What does the display show when a sensor is abnormal?

The LCD shows "Err" when the sensor is abnormal, and the output falls below 2 mA.

Which baud rates can be selected for communication?

The baud rate can be selected from 57.6, 38.4, 28.8, 19.2, 9.6, 4.8 and 2.4 Kbps.

How long is the quality warranty?

Failures caused by product quality problems are covered for 18 months from factory shipment.