The Kedian HCH-2000C ultrasonic thickness gauge user manual provides operation, measurement, storage, and maintenance instructions.
The HCH-2000C ultrasonic thickness gauge is a compact, low‑power, handheld instrument that uses digital electronics and a micro‑computer to measure the thickness of boilers, pressure vessels, pipes, metal parts, glass, plastic, and acrylic. It offers a wide measurement range, high accuracy, and intuitive touch‑button operation for field use.
The manual explains how to power the device, select probes, set sound speed, perform manual storage measurements, read stored results, and maintain the instrument. It also covers various measurement methods, causes of reading errors, and preventive measures to ensure reliable thickness readings. This manual is written in Chinese.
| Measurement range | 0.7‑199.9 mm (45 # steel) |
|---|---|
| Display accuracy | 0.1 mm |
| Measurement error | ±1 % × thickness ± 0.1 mm |
| Measurement mode | Manual storage measurement |
| Storage capacity | 600 measurement points |
| Probe frequency | 5 MHz |
| Sound‑speed range | 1000‑9990 m/s |
| Power supply | 1.2‑1.5 V alkaline or rechargeable battery |
| Dimensions | 124 × 51 × 27 mm³ |
| Weight | 160 g (incl. battery) |
| Operating temperature | -10 °C to 60 °C |
| Automatic shutdown | Power off after 5 min standby |
Place a suitable coupling agent on a standard block, press the CAL key, and the display will show the calibration value (e.g., 4.0). Release CAL and the instrument will display the calibrated value; repeat if the reading is inaccurate.
The device can store up to 600 measurement points.
If the instrument remains in standby for five minutes, it will automatically shut down.