This manual provides setup and calibration procedures for various Ishida industrial and non-system retail scales and balances sold by Rice Lake Weighing…
This document serves as a compilation of setup and calibration procedures for a wide range of Ishida weighing products, including retail scales, industrial scales, and balances. It is intended for use by service technicians responsible for installing and servicing these products to ensure compliance with NTEP requirements and proper operation. The manual covers specific models such as the NOVA-II and BC-3000 for retail applications, and the MTX/MTB, MTE, and MG-3000 series for industrial use.
| Standard calibration weight (MB Series) | 1/2 capacity |
|---|---|
| Standard calibration weight (IB/SB Series) | 1/2 capacity |
| Display tolerance (MB Series) | 0.4% |
| Display tolerance (IB/SB Series) | 1% |
All work performed on the NOVA-II, including setup and calibration, must be done by an authorized scale distributor to ensure compliance with NTEP requirements.
If [o-Err] appears on the display, the weight exceeds the scale capacity.
No, MB Series scales cannot be calibrated; they are stamped or sealed by local weights/measurement offices.