Operators manual for iseg VME STANDARD series high voltage supplies, covering technical data, front panel, handling, and VME interface.
This operators manual covers the iseg VME STANDARD series high voltage supplies, including the VHQ one- and two-channel modules. It describes general information, technical data, VHQ description, front panel, handling, VME interface, a program example, and two appendices.
The modules provide manual control and operation via the VME bus, with output voltages from 0 to 5 kV per channel depending on model, hardware and software ramps, and protection features such as current and voltage limits and INHIBIT.
| Output voltage | 0 ... 2 kV to 0 ... 5 kV depending on model |
|---|---|
| Output current per channel | 0 ... 3 mA to 0 ... 1 mA depending on model |
| Ripple (typical) | < 0.5 mV P-P to < 1 mV P-P depending on model |
| Ripple (maximum) | 2 mV P-P |
| Stability | ΔVO < 5 × 10^-5 (idle to max. load); ΔVO/ΔVINPUT < 5 × 10^-5 |
| Temperature coefficient | < 5 × 10^-5 /K |
| LCD display | 4 digits with sign; voltage in V or current in µA |
| Resolution of current measurement | 1 µA |
| Resolution of voltage measurement | 1 V |
| Accuracy of current measurement | ±(0.05% IO + 0.02% IO max + 1 digit) for one year |
| Accuracy of voltage measurement | ±(0.05% VO + 0.02% VO max + 1 digit) for one year |
| Hardware ramp rate | 500 V/s (on HV-ON/-OFF) |
| Software ramp rate | 2 ... 255 V/s |
| Current trip reaction time | < 60 ms |
| Power requirements | ±12 V (< 850 mA, one channel < 450 mA); +5 V (< 300 mA) |
| Packing | VME #2 / 6U / 164 mm deep |
| Operating temperature | 0 ... +50 °C |
| Storage temperature | -20 ... +60 °C |
Short supervisory access (AM = 0x2D) and short nonprivileged access (AM = 0x29).
With a rotary switch on the cover side; the chosen polarity is shown by an LED and the LCD sign. Changing polarity under power is not allowed.
High voltage is generated at the HV output with a 500 V/s hardware ramp to the voltage set by the 10-turn potentiometer.
Less than 60 ms.